Endoluminal high-resolution MR imaging protocol for colon walls analysis in a mouse model of colitis

MAGMA. 2016 Aug;29(4):657-69. doi: 10.1007/s10334-016-0539-2. Epub 2016 Mar 10.

Abstract

Objective: An endoluminal magnetic resonance (MR) imaging protocol including the design of an endoluminal coil (EC) was defined for high-spatial-resolution MR imaging of mice gastrointestinal walls at 4.7 T.

Materials and methods: A receive-only radiofrequency single-loop coil was developed for mice colon wall imaging. Combined with a specific protocol, the prototype was first characterized in vitro on phantoms and on vegetables. Signal-to-noise ratio (SNR) profiles were compared with a quadrature volume birdcage coil (QVBC). Endoluminal MR imaging protocol combined with the EC was assessed in vivo on mice.

Results: The SNR measured close to the coil is significantly higher (10 times and up to 3 mm of the EC center) than the SNR measured with the QVBC. The gain in SNR can be used to reduce the in-plane pixel size up to 39 × 39 µm(2) (234 µm slice thickness) without time penalty. The different colon wall layers can only be distinguished on images acquired with the EC.

Conclusion: Dedicated EC provides suitable images for the assessment of mice colon wall layers. This proof of concept provides gains in spatial resolution and leads to adequate protocols for the assessment of human colorectal cancer, and can now be used as a new imaging tool for a better understanding of the pathology.

Keywords: Colitis; Colorectal neoplasms; Instrumentation; Magnetic resonance imaging.