A diamond 14 MeV neutron energy spectrometer with high energy resolution

Rev Sci Instrum. 2016 Feb;87(2):023503. doi: 10.1063/1.4940929.

Abstract

A self-standing single-crystal chemical vapor deposited diamond was obtained using lift-off method. It was fabricated into a radiation detector and response function measurements for 14 MeV neutrons were taken at the fusion neutronics source. 1.5% of high energy resolution was obtained by using the (12)C(n, α)(9)Be reaction at an angle of 100° with the deuteron beam line. The intrinsic energy resolution, excluding energy spreading caused by neutron scattering, slowing in the target and circuit noises was 0.79%, which was also the best resolution of the diamond detector ever reported.