Impact of MR-Based Attenuation Correction on Neurologic PET Studies

J Nucl Med. 2016 Jun;57(6):913-7. doi: 10.2967/jnumed.115.164822. Epub 2016 Jan 28.

Abstract

Hybrid PET and MR scanners have become a reality in recent years, with the benefits of reduced radiation exposure, reduction of imaging time, and potential advantages in quantification. Appropriate attenuation correction remains a challenge. Biases in PET activity measurements were demonstrated using the current MR-based attenuation-correction technique. We aimed to investigate the impact of using a standard MR-based attenuation correction technique on the clinical and research utility of a PET/MR hybrid scanner for amyloid imaging.

Methods: Florbetapir scans were obtained for 40 participants on a hybrid scanner with simultaneous MR acquisition. PET images were reconstructed using both MR- and CT-derived attenuation maps. Quantitative analysis was performed for both datasets to assess the impact of MR-based attenuation correction to absolute PET activity measurements as well as target-to-reference ratio (SUVR). Clinical assessment was also performed by a nuclear medicine physician to determine amyloid status based on the criteria in the Food and Drug Administration prescribing information for florbetapir.

Results: MR-based attenuation correction led to underestimation of PET activity for most parts of the brain, with a small overestimation for deep brain regions. There was also an overestimation of SUVRs with cerebellar reference. SUVR measurements obtained from the 2 attenuation-correction methods were strongly correlated. Clinical assessment of amyloid status resulted in identical classification as positive or negative regardless of the attenuation-correction methods.

Conclusion: MR-based attenuation correction causes biases in quantitative measurements. The biases may be accounted for by a linear model, although the spatial variation cannot be easily modeled. The quantitative differences, however, did not affect clinical assessment as positive or negative.

Keywords: PET/MR; amyloid imaging; attenuation correction.

Publication types

  • Research Support, Non-U.S. Gov't
  • Research Support, N.I.H., Extramural

MeSH terms

  • Amyloid / metabolism
  • Brain / diagnostic imaging
  • Brain / metabolism
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging*
  • Multimodal Imaging*
  • Positron-Emission Tomography*

Substances

  • Amyloid