INFLUENCE OF THE IN-PLANE ARTEFACT IN CHEST TOMOSYNTHESIS ON PULMONARY NODULE SIZE MEASUREMENTS

Radiat Prot Dosimetry. 2016 Jun;169(1-4):199-203. doi: 10.1093/rpd/ncv536. Epub 2016 Jan 13.

Abstract

The aim of the present study was to investigate how the in-plane artefact present in the scan direction around structures in tomosynthesis images should be managed when measuring the size of nodules in chest tomosynthesis images in order to achieve acceptable measurement accuracy. Data from measurements, performed by radiologists, of the longest diameter of artificial nodules inserted in chest tomosynthesis images were used. The association between the measurement error and the direction of the longest nodule diameter, relative to the scan direction, was evaluated using the Kendall rank correlation coefficient. All of the radiologists had chosen to not include the artefact in the measurements. Significant association between measurement error and the direction of the longest diameter was found for nodules larger than 12 mm, which indicates that, for these nodules, there is a risk of underestimating the nodule size if the in-plane artefact is omitted from manual diameter measurements.