Concentration Monitoring of Plasma Ribavirin: Validation of a Liquid Chromatography-Mass Spectrometric Method and Clinical Sample Collection

Ther Drug Monit. 2016 Feb;38(1):50-8. doi: 10.1097/FTD.0000000000000232.

Abstract

Background: A liquid chromatography-tandem mass spectrometric (LC-MS/MS) method was developed for routine measurement of ribavirin concentrations in EDTA-anticoagulated plasma.

Methods: After protein precipitation, we used a bridged ethylene hybrid (hydrophilic interaction) chromatography column, 0.1 mmol/L ammonium formate pH 3.0, and a gradient of 85%-96% acetonitrile to achieve baseline separation of ribavirin from isobaric uridine. Quantitation was assured using both primary (m/z 245.3 > 113.0) and secondary transitions (m/z 245.3 > 96.0) of the protonated species. Chromatographic separation and column washing also negated interference from major phospholipid species.

Results: There was a linear relationship between concentration and response to 10 mg/L, with a minimum detectable level and a minimum level of quantitation both of 0.1 mg/L. Imprecision within the assay was <10% at 0.1 mg/L and <6% between assays for concentrations >0.4 mg/L. Bias was <4%. In clinical samples (n = 12), there was no difference in ribavirin concentrations obtained by an established liquid chromatographic assay with ultraviolet detection. Ribavirin concentrations were stable in plasma stored at room temperature for 3 days but then decreased significantly on day 7. Plasma concentrations were stable for 15 weeks at -20 °C. Concentrations in plasma separated from whole blood at room temperature fell by a median of 19.4% at 4 hours and then rose substantially (median 251% by 3 days). Dose-normalized ribavirin concentrations reached a steady state after a mean of >6 weeks treatment in 76 patients with hepatitis C.

Conclusions: A hydrophilic interaction liquid chromatography-tandem mass spectrometric method to measure ribavirin in plasma was developed. Samples for ribavirin estimation should be kept at 4 °C, separated within 2 hours of collection and stored at 4 °C before analysis, with long-term storage at -20 °C. This method was applied to a study of the ribavirin therapeutic monitoring in patients with hepatitis C.

Publication types

  • Validation Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Antiviral Agents / blood
  • Child
  • Child, Preschool
  • Chromatography, Liquid / methods*
  • Drug Monitoring / methods*
  • Drug Stability
  • Drug Storage
  • Edetic Acid / chemistry
  • Hepatitis C / drug therapy
  • Humans
  • Middle Aged
  • Ribavirin / blood*
  • Tandem Mass Spectrometry / methods*
  • Time Factors
  • Young Adult

Substances

  • Antiviral Agents
  • Ribavirin
  • Edetic Acid