CT angiography of the aorta using 80 kVp in combination with sinogram-affirmed iterative reconstruction and automated tube current modulation: Effects on image quality and radiation dose

J Med Imaging Radiat Oncol. 2016 Apr;60(2):187-93. doi: 10.1111/1754-9485.12425. Epub 2016 Jan 7.

Abstract

Introduction: The objective of this study was to evaluate image quality and radiation dose of a CT angiography (CTA) protocol using 80 kVp in combination with iterative reconstruction and automated tube current modulation.

Methods: Ninety-five aortic CTA examinations were included in this study. A novel 80 kVp aortic CTA-protocol with iterative reconstruction was introduced in our department in March 2012 for patients with a body mass index (BMI) below 32 kg/m(2). The first 72 consecutive examinations were retrospectively assigned to group A (56 patients, 42 men, 14 women, mean age 69.6 ± 10.7 years, BMI range 19.7-31.1 kg/m(2)). For comparison, the last 23 consecutive examinations performed with the old protocol (100 kVp) were assigned to group B (21 patients, 13 men, 8 women, mean age 67.4 ± 11.1 years, BMI range 19.7-31.9 kg/m(2)). Thoracic and abdominal contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR) and aortic attenuation were assessed. Subjective image quality was rated on a 5-point scale (1 = non diagnostic; 5 = excellent). Furthermore, dose length product (DLP) and volumetric computed tomography dose index (CTDIvol) were analysed.

Results: All examinations achieved diagnostic image quality. Attenuation of the aorta was significantly higher in group A compared with B (thoracic: 443.5 ± 90.5 Hounsfield units (HU) vs. 296.0 ± 61.0 HU; abdominal: 426.3 ± 94.2 HU vs. 283.6 ± 60.5 HU; P < 0.05, respectively). CNR, SNR and subjective image quality were comparable between both groups (CNR: 12.8 ± 3.7 vs. 13.0 ± 7.4; SNR 14.4 ± 3.9 vs. 14.9 ± 8.2; subjective image quality: 4.3 ± 0.6 vs. 4.5 ± 0.6; P > 0.05, respectively). CTDIvol and DLP were significantly lower in group A (1.9 ± 0.5 mGy; 139.2 ± 41.1 mGy × cm) as compared with group B (4.2 ± 1.4 mGy; 292.1 ± 91.5 mGy × cm; P < 0.001, respectively).

Conclusion: Low-dose CTA of the aorta using 80 kVp with iterative reconstruction enables a significant dose reduction of up to 50% compared with a 100 kVp protocol in patients with a BMI below 32 kg/m(2) while diagnostic image quality is maintained.

Keywords: aorta; computed tomography; diagnostic imaging; radiation dose; vascular.

MeSH terms

  • Aged
  • Algorithms
  • Aortic Diseases / diagnostic imaging*
  • Aortography / instrumentation
  • Aortography / methods*
  • Computed Tomography Angiography / instrumentation
  • Computed Tomography Angiography / methods*
  • Feedback
  • Female
  • Humans
  • Male
  • Radiation Dosage*
  • Radiation Exposure / analysis*
  • Radiographic Image Enhancement / methods
  • Radiographic Image Interpretation, Computer-Assisted / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity