[Electrochemical behavior of 3 magnetic alloys used for prosthetic retention]

Minerva Stomatol. 1989 Jun;38(6):605-9.
[Article in Italian]

Abstract

By means of potentiodynamic techniques the electrochemical behaviour of a ferromagnetic alloy Pd-Co and three types of caps (AISI 316 stainless steel, Pd, Ti) for permanent Sm-Co magnets has been investigated in artificial saliva and in Ringer's solution. In addition short-circuit measurements were performed on several couples Pd-Co/AISI 316, Ti, Pd caps. From our findings a good corrosion resistance of all the materials is evidenced in artificial saliva. In Ringer's solution, which is more aggressive, Pd-Co shows no tendency to passivity.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cobalt
  • Corrosion
  • Dental Alloys*
  • Denture Retention*
  • Electrochemistry
  • Ferric Compounds*
  • Humans
  • Isotonic Solutions
  • Palladium
  • Ringer's Solution

Substances

  • Dental Alloys
  • Ferric Compounds
  • Isotonic Solutions
  • ferrite
  • Cobalt
  • Palladium
  • Ringer's Solution