Measurement of (124)I

Appl Radiat Isot. 2016 Mar:109:349-353. doi: 10.1016/j.apradiso.2015.12.002. Epub 2015 Dec 4.

Abstract

The paper describes the measurements performed at IFIN-HH regarding the creation of a Romanian (124)I standard, consisting of: absolute standardization of the solution by the application of the 4πβ(PC)-γ coincidence method; Calibration of the CENTRONIC IG12/20A ionization chamber with a standardised solution and comparison with a calculated efficiency; γ-ray spectrometry activity measurement and determination of the impurity levels; Comparison of the results of the three methods.

Keywords: (124)I standardization; 4πβ(PC)-γ coincidence method; Ionization chamber; γ-ray spectrometry.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Calibration / standards*
  • Equipment Design
  • Equipment Failure Analysis
  • Iodine Radioisotopes / analysis*
  • Iodine Radioisotopes / standards*
  • Radiation Dosimeters / standards*
  • Reference Values
  • Reproducibility of Results
  • Romania
  • Sensitivity and Specificity
  • Spectrometry, Gamma / instrumentation*
  • Spectrometry, Gamma / standards*

Substances

  • Iodine Radioisotopes