Assessing sample attenuation parameters for use in low-energy efficiency transfer in gamma-ray spectrometry

Appl Radiat Isot. 2016 Mar:109:547-550. doi: 10.1016/j.apradiso.2015.11.077. Epub 2015 Dec 2.

Abstract

We present a numerical fitting method for transmission data that outputs an equivalent sample composition. This output is used as input to a generalised efficiency transfer model based on the EFFTRAN software integrated in a LIMS. The procedural concept allows choosing between efficiency transfer with a predefined sample composition or with an experimentally determined composition based on a transmission measurement. The method can be used for simultaneous quantification of low-energy gamma emitters like (210)Pb, (241)Am, (234)Th in typical environmental samples.

Keywords: Attenuation coefficients; Attenuation correction; EFFTRAN; Efficiency transfer; Gamma-ray self-attenuation; Gamma-ray spectrometry; Transmission measurement.