A beta-gamma coincidence system has been developed for measuring (198)Au activity in gold foils. The system was validated by Monte Carlo simulations and by measuring the activity of a (60)Co point-source. To study effects such as self-shielding of beta particles in gold foils, (198)Au activity measurements and simulations were performed for various scintillators and foil sizes. The measured (198)Au activities were ~1% above the reference activity, which might be due to self-shielding of beta particles. The measured and simulated (198)Au activities agreed, suggesting feasibility of precise activity measurement.
Keywords: (198)Au; Beta–gamma coincidence counting; Geant4; Thermal neutron field.
Copyright © 2015 Elsevier Ltd. All rights reserved.