Simple Pretreatment and HILIC Separation for LC-ESI-MS/MS Determination of Adenosine in Human Plasma

Anal Sci. 2015;31(11):1189-92. doi: 10.2116/analsci.31.1189.

Abstract

A simple pretreatment method and separation mode for the LC-ESI-MS/MS determination of adenosine in human plasma have been developed. Deproteinization by acetonitrile and ultrafiltration followed by chromatographic separation with a hydrophilic interaction chromatographic (HILIC) column give a highly sensitive MS/MS response without ionic suppression caused by the matrix compounds in human plasma. In addition, the presence of ammonium acetate in the mobile phase contributes to high sensitivity in MS/MS detection, facilitating the ionization of adenosine. This method seems to be amenable to the treatment of many samples in clinical practice.

MeSH terms

  • Adenosine / analysis*
  • Adenosine / isolation & purification*
  • Analytic Sample Preparation Methods / methods*
  • Blood Chemical Analysis / methods*
  • Calibration
  • Chromatography, Liquid / methods*
  • Humans
  • Hydrophobic and Hydrophilic Interactions*
  • Spectrometry, Mass, Electrospray Ionization / methods*
  • Tandem Mass Spectrometry
  • Ultrafiltration