Fast ion energy distribution from third harmonic radio frequency heating measured with a single crystal diamond detector at the Joint European Torus

Rev Sci Instrum. 2015 Oct;86(10):103501. doi: 10.1063/1.4931755.

Abstract

Neutron spectroscopy measurements with a single crystal diamond detector have been carried out at JET, for the first time in an experiment aimed at accelerating deuterons to MeV energies with radio frequency heating at the third harmonic. Data are interpreted by means of the expected response function of the detector and are used to extract parameters of the highly non-Maxwellian distribution function generated in this scenario. A comparison with observations using a time of flight and liquid scintillator neutron spectrometers is also presented. The results demonstrate the capability of diamond detectors to contribute to fast ion physics studies at JET and are of more general relevance in view of the application of such detectors for spectroscopy measurements in the neutron camera of next step tokamak devices.