We report the microsurgical rescue and removal of a Pipeline stent embolization of a giant internal carotid artery terminus aneurysm. After the initial placement of a Pipeline Embolization Device (PED), it migrated proximally to the cavernous carotid with the distal end free in the middle of the aneurysm, resulting in only partial aneurysm neck coverage. The patient underwent microsurgical rescue with trapping, bypass, and opening of the aneurysm with PED removal. The vessel remained patent in the proximal segment previously covered by the Pipeline stent. Microsurgical rescue for definitive aneurysm treatment with PED removal can be safe and effective for aneurysms unsuccessfully treated with PED.
Keywords: Endovascular treatment; Giant cerebral aneurysm; Open vascular surgery; Pipeline Embolization Device; Rescue microsurgery.