Measuring skin dose in CT examinations under complex geometries: Instruments, methods and considerations

Phys Med. 2015 Dec;31(8):1005-1014. doi: 10.1016/j.ejmp.2015.08.001. Epub 2015 Sep 26.

Abstract

Objective: To investigate skin dose in Computed Tomography (CT) and its dependence on scanning geometry.

Materials and methods: Measurements of entrance surface air kerma (ESAK) in free air and entrance skin dose (ESD) on an anthropomorphic phantom were performed in a 64-slice CT scanner, using two different instruments: the Dose Profiler (DP) and the QED skin diode (QEDSD). Using DP and QEDSD, the ESAK rate profiles at the isocenter and at different distances from it, were measured using axial scans. Using DP and helical scans the ESAK rate profile in the Z-axis was acquired. The same profile was acquired with the QEDSD also, using many axial scans and manual table translation. ESD measurements were performed with the DP and QEDSD, in axial and helical scan mode.

Results: ESAK measurements with DP and QEDSD were in good agreement, for both point dose and profile measurements. The agreement was also good for ESD measurements but not for helical scans, due to variable X-ray beam overlapping and different tube angular positions at each scan start. It was observed that the ESD values at different Y-axis offsets were comparable to the respective ESAK values recorded at the same Y-axis offset distances without the phantom.

Conclusions: Both DP and QEDSD were proven suitable for performing point ESD measurements. However, calculating the skin dose distribution in CT examinations is a very challenging task. A practical approach would be for CT scanners to provide a conservative estimate of the peak skin dose using the isocenter ESAK value.

Keywords: CT; ESD; Interventional; Skin dose.