Master slave en-face OCT/SLO

Biomed Opt Express. 2015 Aug 27;6(9):3655-69. doi: 10.1364/BOE.6.003655. eCollection 2015 Sep 1.

Abstract

Master Slave optical coherence tomography (MS-OCT) is an OCT method that does not require resampling of data and can be used to deliver en-face images from several depths simultaneously. As the MS-OCT method requires important computational resources, the number of multiple depth en-face images that can be produced in real-time is limited. Here, we demonstrate progress in taking advantage of the parallel processing feature of the MS-OCT technology. Harnessing the capabilities of graphics processing units (GPU)s, information from 384 depth positions is acquired in one raster with real time display of up to 40 en-face OCT images. These exhibit comparable resolution and sensitivity to the images produced using the conventional Fourier domain based method. The GPU facilitates versatile real time selection of parameters, such as the depth positions of the 40 images out of the set of 384 depth locations, as well as their axial resolution. In each updated displayed frame, in parallel with the 40 en-face OCT images, a scanning laser ophthalmoscopy (SLO) lookalike image is presented together with two B-scan OCT images oriented along rectangular directions. The thickness of the SLO lookalike image is dynamically determined by the choice of number of en-face OCT images displayed in the frame and the choice of differential axial distance between them.

Keywords: (110.4500) Optical coherence tomography; (120.3890) Medical optics instrumentation; (170.0110) Imaging systems; (200.4960) Parallel processing; (330.4460) Ophthalmic optics and devices.