Characterisation of a fibre optic Raman probe within a hypodermic needle

Anal Bioanal Chem. 2015 Nov;407(27):8311-20. doi: 10.1007/s00216-015-9021-7. Epub 2015 Sep 28.

Abstract

We demonstrate the first use of a multifibre Raman probe that fits inside the bore of a hypodermic needle. A Raman probe containing multiple collection fibres provides improved signal collection efficiency in biological samples compared with a previous two-fibre design. Furthermore, probe performance (signal-to-noise ratios) compared favourably with the performance achieved in previous Raman microscope experiments able to distinguish between benign lymph nodes, primary malignancies in lymph nodes and secondary malignancies in lymph nodes. The experimental measurements presented here give an indication of the sampling volume of the Raman needle probe in lymphoid tissues. Liquid tissue phantoms were used that contained scattering medium encompassing a range of scattering properties similar to those of a variety of tissue types, including lymph node tissues. To validate the appropriateness of the phantoms, the sampling depth of the probe was also measured in excised lymph node tissue. More than 50 % of Raman photons collected were found to originate from between the tip of the needle and a depth of 500 μm into the tissue. The needle probe presented here achieves spectral quality comparable to that in numerous studies previously demonstrating Raman disease discrimination. It is expected that this approach could achieve targeted subcutaneous tissue measurements and be viable for use for the in vivo Raman diagnostics of solid organs located within a few centimetres below the skin's surface. Graphical Abstract Schematic of multi-fibre Raman needle probe with disposible tips and proximal optical filtration.

Keywords: Diagnostics; Lymph nodes; Needle; Non-invasive; Raman spectroscopy.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Equipment Design
  • Fiber Optic Technology / instrumentation*
  • Humans
  • Lymph Nodes / pathology*
  • Lymphatic Metastasis / diagnosis*
  • Lymphatic Metastasis / pathology
  • Needles*
  • Neoplasms / diagnosis*
  • Neoplasms / pathology
  • Spectrum Analysis, Raman / instrumentation*