A compact gas chromatograph and pre-column concentration system for enhanced in-field separation of levoglucosan and other polar organic compounds

J Chromatogr A. 2015 Oct 23:1417:73-8. doi: 10.1016/j.chroma.2015.09.015. Epub 2015 Sep 8.

Abstract

Portable and compact instruments for separating and detecting organic compounds are needed in the field for environmental studies. This is especially the case for pollution studies as in-field detection of organic compounds helps identify sources of pollution. Here we describe the development of a compact GC and simple pre-concentrator coupled to a MS detector. This simple system can easily be incorporated into portable instrumentation. Combining the pre-concentrator and compact column has the advantage of decoupling separation from manual injection and enhances separation of environmentally relevant polar organic compounds, such as levoglucosan. A detection limit of 2.2 ng was obtained for levoglucosan. This simple design has the potential to expand the use of gas chromatography as a routine in-field separation technique.

Keywords: Compact GC; Field sampling; Levoglucosan; Pre-concentration.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Air Pollutants / isolation & purification*
  • Chromatography, Gas / instrumentation
  • Chromatography, Gas / methods*
  • Glucose / analogs & derivatives*
  • Glucose / isolation & purification
  • Hydrocarbons / isolation & purification*
  • Polycyclic Aromatic Hydrocarbons / isolation & purification

Substances

  • Air Pollutants
  • Hydrocarbons
  • Polycyclic Aromatic Hydrocarbons
  • 1,6-anhydro-beta-glucopyranose
  • Glucose