Static imaging of the electrical impedance tomography on cylinder physical phantom

Biomed Mater Eng. 2015:26 Suppl 1:S1381-8. doi: 10.3233/BME-151436.

Abstract

Static imaging of the electrical impedance tomography can obtain the absolute electrical conductivity distribution at one section of the subject. The test is performed on a cylinder physical phantom in which slim rectangle, hollow cylinder, small rectangle or three cylinders are selected to simulate complex conductivity perturbation objects. The measurement data is obtained by a data acquisition system with 32 compound electrodes. A group of static images of conductivity distribution in the cylinder phantom are reconstructed by the modified Newton-Raphson algorithm with two kinds of regularization methods. The results show correct position, size, conductivity difference, and similar shape of the perturbation objects in the images.

Keywords: Electrical impedance; physical phantom; reconstruction algorithm; static imaging; tomography.