Determination of reduced homocysteine in human serum by elemental labelling and liquid chromatography with ICP-MS and ESI-MS detection

Anal Bioanal Chem. 2015 Oct;407(26):7899-906. doi: 10.1007/s00216-015-8956-z. Epub 2015 Sep 11.

Abstract

Analytical methods allowing sensitive determination of reduced homocysteine (rHcy), one of the so-called biothiols, in human serum is a topic of growing interest due to its close relation to several human disorders, mainly cardiovascular diseases. Although most widely used analytical strategies to determine total Hcy involve derivatization by means of fluorescent labels, this work proposes the use of ebselen, a Se-containing labelling agent to derivatize the reactive sulfhydryl group of the Hcy molecule in its "free" reduced form, which is more likely to play different roles in disease pathogenesis. Optimization of the derivatization and separation conditions by high-performance liquid chromatography (HPLC) to isolate the excess of derivatizing reagent is carried out here using UV/VIS detection. Further, the study of the Se labelling reaction by electrospray ionization tandem mass spectrometry (ESI-MS/MS) provides a stoichiometry of the derivative of 1:1. The main advantage of using ebselen as a labelling agent is the presence of the Se atom in the molecule that allows the use of inductively coupled plasma mass spectrometry (ICP-MS) as a sensitive and selective Se detector. The coupling of HPLC with ICP-MS provided excellent features for the determination of Se-derivatized rHcy (detection limit of 9.6 nM) in real samples. Quantification was accomplished by using post-column isotope dilution (ID) of Se in serum samples, after precipitation of the main serum proteins. Quantitative results for "free" rHcy turned out to be around 0.18-0.22 μM in serum samples from healthy individuals that could be directly analyzed without sample preconcentration.

Keywords: Ebselen; Isotope dilution; LC–ICP-MS; Labelling; Reduced homocysteine.

Publication types

  • Research Support, Non-U.S. Gov't
  • Validation Study

MeSH terms

  • Chromatography, High Pressure Liquid / methods*
  • Homocysteine / analysis
  • Homocysteine / blood*
  • Humans
  • Limit of Detection
  • Oxidation-Reduction
  • Spectrometry, Mass, Electrospray Ionization / methods*
  • Tandem Mass Spectrometry / methods

Substances

  • Homocysteine