Various verifications were performed to apply JENDL-4.0 as nuclear data for a newly developed treatment planning system with a homogeneous or precise human-like phantom. The nitrogen dose calculated by JENDL-4.0 differed slightly from that calculated by ENDF/B-VII.0. However, the total weighted dose-based dose volume histogram in the boron neutron capture therapy (BNCT) treatment for brain tumors calculated by JENDL-4.0 was in good agreement with the results of the ENDF/B-VII.0 calculation. Therefore, calculation with JENDL-4.0 can be applied to the BNCT dose calculation.
Keywords: BNCT; Computational dosimetry; Monte Carlo; Nuclear data; Treatment planning system.