Determination of the Density of Protein Particles Using a Suspended Microchannel Resonator

J Pharm Sci. 2015 Dec;104(12):4034-4040. doi: 10.1002/jps.24635. Epub 2015 Sep 7.

Abstract

One of the analytical tools for characterization of subvisible particles, which gained popularity over the last years because of its unique capabilities, is the resonance mass measurement technique. However, a challenge that this technique presents is the need to know the exact density of the measured particles in order to obtain accurate size calculations. The density of proteinaceous subvisible particles has not been measured experimentally yet and to date researchers have been using estimated density values. In this paper, we report for a first-time experimental measurements of the density of protein particles (0.2-5 μm in size) using particles created by stressing three different proteins using four different types of stress conditions. Interestingly, the particle density values that were measured varied between 1.28 and 1.33 g/cm(3) and were lower than previous estimates. Furthermore, it was found that although the density of proteinaceous particles was affected to a very low degree by the stress conditions used to generate them, there is relatively larger difference between particles originating from different classes of proteins (e.g., monoclonal antibody vs. bovine serum albumin).

Keywords: Archimedes; density determination; microparticles; nanoparticles; particle size; protein aggregation; proteins; resonance mass measurement; suspended microchannel resonator.

MeSH terms

  • Light
  • Particle Size
  • Proteins / chemistry*

Substances

  • Proteins