Background: An improved version of the bioMérieux NucliSENS(®) EasyQ(®) HIV-1 v2.0 has been introduced to overcome the underquantification observed with previous versions, especially with non-B HIV-1 subtypes.
Objectives: Comparing bioMérieux NucliSENS(®) EasyQ(®) HIV-1 v2.0 versus Roche Cobas CA/CTM v2.0 and Abbott RealTime HIV-1 assays for HIV-1 group M and non-M (N, O, P) viral load measurement.
Study design: The three assays were tested in parallel on 103HIV-1 group M plasma samples, and on non-group M HIV-1 culture supernatants.
Results: Values obtained for the 103HIV-1 group M plasma samples tested with bioMérieux assay showed good overall correlation compared to the 2 others. The Roche Cobas assay gave higher values than the bioMérieux assay, while Abbott and bioMérieux both displayed similar results. However, analysis showed a wider dispersion in results when comparing the bioMérieux NucliSENS(®) EasyQ(®) HIV-1 v2.0 and the other 2 techniques. All data taken into account, we observed frequent discrepancies in quantification, of the plasma samples and major differences above 1log in 10/72 (13.8%). The quantification of non-M HIV groups in culture supernatant has shown variable results, with better quantification of HIV-O and of HIV-N respectively with the Abbott assay and the bioMérieux assay.
Conclusions: The bioMérieux NucliSENS(®) EasyQ(®) HIV-1 v2.0 showed improved sensitivity to non-B HIV-M subtypes compared to previous versions. Notwithstanding, we observed frequent discrepancies and a wide dispersion in results when comparing bioMérieux NucliSENS(®) EasyQ(®) HIV-1 v2.0 and the other 2 techniques.
Keywords: Genetic diversity; HIV-1; HIV-O; Quantification; Viral load.