Transference of CALIPER pediatric reference intervals to biochemical assays on the Roche cobas 6000 and the Roche Modular P

Clin Biochem. 2016 Jan;49(1-2):139-49. doi: 10.1016/j.clinbiochem.2015.08.018. Epub 2015 Aug 19.

Abstract

Objectives: The Canadian Laboratory Initiative on Pediatric Reference Intervals (CALIPER) has recently established pediatric age- and sex-specific reference intervals for over 85 biochemical markers on the Abbott Architect system. Previously, CALIPER reference intervals for several biochemical markers were successfully transferred from Abbott assays to Roche, Beckman, Ortho, and Siemens assays. This study further broadens the CALIPER database by performing transference and verification for 52 biochemical assays on the Roche cobas 6000 and the Roche Modular P.

Design and methods: Using CLSI C28-A3 and EP9-A2 guidelines, transference of the CALIPER reference intervals was attempted for 16 assays on the Roche cobas 6000 and 36 on the Modular P. Calculated reference intervals were further verified using 100 healthy CALIPER samples.

Results: Most assays showed strong correlation between assay systems and were transferable from Abbott to the Roche cobas 6000 (81%) and the Modular P (86%). Bicarbonate and magnesium were not transferable on either system and calcium and prealbumin were not transferable to the Modular P. Of the transferable analytes, 62% and 61% were verified on the cobas 6000 and the Modular P, respectively.

Conclusions: This study extends the utility of the CALIPER database to two additional analytical systems, which facilitates the broad application of CALIPER reference intervals at pediatric centers utilizing Roche biochemical assays. Transference studies across different analytical platforms can later be collectively analyzed in an attempt to develop common reference intervals across all clinical chemistry instruments to harmonize laboratory test interpretation in diagnosis and monitoring of pediatric disease.

Keywords: Abbott Architect; CALIPER; Pediatrics; Reference intervals; Roche Modular P; Roche cobas 6000.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Chemistry, Clinical / instrumentation*
  • Child
  • Child, Preschool
  • Humans
  • Infant
  • Pediatrics*
  • Reference Values
  • Young Adult