"Unfocus" on foc.us: commercial tDCS headset impairs working memory

Exp Brain Res. 2016 Mar;234(3):637-43. doi: 10.1007/s00221-015-4391-9. Epub 2015 Aug 18.

Abstract

In this study, we tested whether the commercial transcranial direct current stimulation (tDCS) headset foc.us improves cognitive performance, as advertised in the media. A single-blind, sham-controlled, within-subject design was used to assess the effect of online and off-line foc.us tDCS-applied over the prefrontal cortex in healthy young volunteers (n = 24) on working memory (WM) updating and monitoring. WM updating and monitoring, as assessed by means of the N-back task, is a cognitive-control process that has been shown to benefit from interventions with CE-certified tDCS devices. For both online and off-line stimulation protocols, results showed that active stimulation with foc.us, compared to sham stimulation, significantly decreased accuracy performance in a well-established task tapping WM updating and monitoring. These results provide evidence for the important role of the scientific community in validating and testing far-reaching claims made by the brain training industry.

Keywords: N-back; Updating; Working memory; foc.us; tDCS.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Cross-Over Studies
  • Female
  • Humans
  • Male
  • Memory Disorders / diagnosis
  • Memory Disorders / etiology
  • Memory, Short-Term / physiology*
  • Prefrontal Cortex / physiology*
  • Psychomotor Performance / physiology*
  • Single-Blind Method
  • Transcranial Direct Current Stimulation / adverse effects
  • Transcranial Direct Current Stimulation / methods*
  • Young Adult