Efficacy of a Bio-Absorbable Antibacterial Envelope to Prevent Cardiac Implantable Electronic Device Infections in High-Risk Subjects

J Cardiovasc Electrophysiol. 2015 Oct;26(10):1111-6. doi: 10.1111/jce.12768. Epub 2015 Sep 6.

Abstract

Introduction: Cardiac implantable electronic device (CIED) infections are potentially preventable complications associated with high morbidity, mortality, and cost. A recently developed bio-absorbable antibacterial envelope (TYRX™-A) might prevent CIED infections in high-risk subjects. However, data regarding safety and efficacy have not been published.

Methods and results: In a single-center retrospective cohort study, we compared the prevalence of CIED infections among subjects with ≥2 risk factors treated with the TYRX™-A envelope (N = 135), the nonabsorbable TYRX™ envelope (N = 353), and controls who did not receive an envelope (N = 636). Infection was ascertained by individual chart review. The mean (95% confidence interval) number of risk factors was 3.08 (2.84-3.32) for TYRX™-A, 3.20 (3.07-3.34) for TYRX™, and 3.09 (2.99-3.20) for controls, P = 0.3. After a minimum 300 days follow-up, the prevalence of CIED infection was 0 (0%) for TYRX™-A, 1 (0.3%) for TYRX™, and 20 (3.1%) for controls (P = 1 for TYRX™-A vs. TYRX™, P = 0.03 for TYRX™-A vs. controls, and P = 0.002 for TYRX™ vs. controls). In a propensity score-matched cohort of 316 recipients of either envelope and 316 controls, the prevalence of infection was 0 (0%) and 9 (2.8%), respectively, P = 0.004. When limited to 122 TYRX™-A recipients and 122 propensity-matched controls, the prevalence of CIED infections was 0 (0%) and 5 (4.1%), respectively, P = 0.024.

Conclusions: Among high-risk subjects, the TYRX™-A bio-absorbable envelope was associated with a very low prevalence of CIED related infections that was comparable to that seen with the nonabsorbable envelope.

Keywords: antibacterial envelope; cardiac implantable electronic device; infection.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Absorbable Implants / statistics & numerical data
  • Aged
  • Anti-Bacterial Agents / administration & dosage*
  • Causality
  • Cohort Studies
  • Defibrillators, Implantable / statistics & numerical data*
  • Delayed-Action Preparations / administration & dosage
  • Drug Implants / administration & dosage*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Pacemaker, Artificial / statistics & numerical data*
  • Prevalence
  • Prosthesis-Related Infections / diagnosis
  • Prosthesis-Related Infections / epidemiology*
  • Prosthesis-Related Infections / prevention & control*
  • Retrospective Studies
  • Risk Factors
  • Sex Distribution
  • Survival Rate
  • Tennessee / epidemiology
  • Treatment Outcome

Substances

  • Anti-Bacterial Agents
  • Delayed-Action Preparations
  • Drug Implants