Radioisotopic Purity of Sodium Pertechnetate 99mTc Produced with a Medium-Energy Cyclotron: Implications for Internal Radiation Dose, Image Quality, and Release Specifications

J Nucl Med. 2015 Oct;56(10):1600-8. doi: 10.2967/jnumed.115.156398. Epub 2015 Jul 23.

Abstract

Cyclotron production of 99mTc is a promising route to supply 99mTc radiopharmaceuticals. Higher 99mTc yields can be obtained with medium-energy cyclotrons in comparison to those dedicated to PET isotope production. To take advantage of this capability, evaluation of the radioisotopic purity of 99mTc produced at medium energy (20-24 MeV) and its impact on image quality and dosimetry was required.

Methods: Thick 100Mo (99.03% and 99.815%) targets were irradiated with incident energies of 20, 22, and 24 MeV for 2 or 6 h. The targets were processed to recover an effective thickness corresponding to approximately 5-MeV energy loss, and the resulting sodium pertechnetate 99mTc was assayed for chemical, radiochemical, and radionuclidic purity. Radioisotopic content in final formulation was quantified using γ-ray spectrometry. The internal radiation dose for 99mTc-pertechnetate was calculated on the basis of experimentally measured values and biokinetic data in humans. Planar and SPECT imaging were performed using thin capillary and water-filled Jaszczak phantoms.

Results: Extracted sodium pertechnetate 99mTc met all provisional quality standards. The formulated solution for injection had a pH of 5.0-5.5, contained greater than 98% of radioactivity in the form of pertechnetate ion, and was stable for at least 24 h after formulation. Radioisotopic purity of 99mTc produced with 99.03% enriched 100Mo was greater than 99.0% decay corrected to the end of bombardment (EOB). The radioisotopic purity of 99mTc produced with 99.815% enriched 100Mo was 99.98% or greater (decay corrected to the EOB). The estimated dose increase relative to 99mTc without any radionuclidic impurities was below 10% for sodium pertechnetate 99mTc produced from 99.03% 100Mo if injected up to 6 h after the EOB. For 99.815% 100Mo, the increase in effective dose was less than 2% at 6 h after the EOB and less than 4% at 15 h after the EOB when the target was irradiated at an incident energy of 24 MeV. Image spatial resolution and contrast with cyclotron-produced 99mTc were equivalent to those obtained with 99mTc eluted from a conventional generator.

Conclusion: Clinical-grade sodium pertechnetate 99mTc was produced with a cyclotron at medium energies. Quality control procedures and release specifications were drafted as part of a clinical trial application that received approval from Health Canada. The results of this work are intended to contribute to establishing a regulatory framework for using cyclotron-produced 99mTc in routine clinical practice.

Keywords: 99mTc-pertechnetate; cyclotron; dosimetry; imaging; radionuclidic and radioisotopic purity.

MeSH terms

  • Cyclotrons*
  • Drug Contamination
  • Isotopes
  • Molybdenum
  • Phantoms, Imaging
  • Positron-Emission Tomography
  • Quality Control
  • Radiation Dosage
  • Radiochemistry / methods*
  • Radiopharmaceuticals / chemistry*
  • Radiopharmaceuticals / pharmacokinetics
  • Sodium Pertechnetate Tc 99m / chemistry
  • Sodium Pertechnetate Tc 99m / isolation & purification*
  • Sodium Pertechnetate Tc 99m / pharmacokinetics

Substances

  • Isotopes
  • Radiopharmaceuticals
  • Molybdenum
  • Sodium Pertechnetate Tc 99m