Non-contact scanning diffuse correlation tomography system for three-dimensional blood flow imaging in a murine bone graft model

Biomed Opt Express. 2015 Jun 26;6(7):2695-712. doi: 10.1364/BOE.6.002695. eCollection 2015 Jul 1.

Abstract

A non-contact galvanometer-based optical scanning system for diffuse correlation tomography was developed for monitoring bone graft healing in a murine femur model. A linear image reconstruction algorithm for diffuse correlation tomography was tested using finite-element method based simulated data and experimental data from a femur or a tube suspended in a homogeneous liquid phantom. Finally, the non-contact system was utilized to monitor in vivo blood flow changes prior to and one week after bone graft transplantation within murine femurs. Localized blood flow changes were observed in three mice, demonstrating a potential for quantification of longitudinal blood flow associated with bone graft healing.

Keywords: (170.0110) Imaging systems; (170.3660) Light propagation in tissues; (170.6960) Tomography; (170.7050) Turbid media.