Simple method for reducing the first-order optical nonlinearity in a heterodyne laser interferometer

Appl Opt. 2015 Jul 10;54(20):6321-6. doi: 10.1364/AO.54.006321.

Abstract

A simple method was proposed by using a tunable attenuator fitted in the reference or measurement arm of a heterodyne laser interferometer to adjust the values of mixing laser beams while the spectrum of the measurement signal is monitored using a signal analyzer. The effectiveness of the proposed method in reducing the first-order optical nonlinearity was verified through experiments. Results indicated that the peak value of the first-order optical nonlinearity could be reduced from 5.15 to 0.24 nm. It was therefore concluded that the proposed method was applicable to ultraprecision laser interferometry.