Diffusion MRI with concurrent magnetic field monitoring

Magn Reson Med. 2015 Oct;74(4):925-33. doi: 10.1002/mrm.25827. Epub 2015 Jul 17.

Abstract

Purpose: Diffusion MRI is compromised by unknown field perturbation during image encoding. The purpose of this study was to address this problem using the recently described approach of concurrent magnetic field monitoring.

Methods: Magnetic field dynamics were monitored during the echo planar imaging readout of a common diffusion-weighted MRI sequence using an integrated magnetic field camera setup. The image encoding including encoding changes over the duration of entire scans were quantified and analyzed. Field perturbations were corrected by accounting for them in generalized image reconstruction. The impact on image quality along with geometrical congruence among different diffusion-weighted images was assessed both qualitatively and quantitatively.

Results: The most significant field perturbations were found to be related to higher-order eddy currents from diffusion-weighting gradients and B0 field drift as well as gradual changes of short-term eddy current behavior and mechanical oscillations during the scan. All artifacts relating to dynamic field perturbations were eliminated by incorporating the measured encoding in image reconstruction.

Conclusion: Concurrent field monitoring combined with generalized reconstruction enhances depiction fidelity in diffusion imaging. In addition to artifact reduction, it improves geometric congruence and thus facilitates image combination for quantitative diffusion analysis.

Keywords: DTI; DWI; artifact correction; diffusion; distortion correction; field monitoring.

MeSH terms

  • Adult
  • Algorithms
  • Artifacts
  • Brain / anatomy & histology
  • Diffusion Magnetic Resonance Imaging / methods*
  • Diffusion Tensor Imaging / methods*
  • Female
  • Humans
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Fields
  • Phantoms, Imaging
  • Temperature
  • Young Adult