Hemodynamics in Idealized Stented Coronary Arteries: Important Stent Design Considerations

Ann Biomed Eng. 2016 Feb;44(2):315-29. doi: 10.1007/s10439-015-1387-3. Epub 2015 Jul 16.

Abstract

Stent induced hemodynamic changes in the coronary arteries are associated with higher risk of adverse clinical outcome. The purpose of this study was to evaluate the impact of stent design on wall shear stress (WSS), time average WSS, and WSS gradient (WSSG), in idealized stent geometries using computational fluid dynamics. Strut spacing, thickness, luminal protrusion, and malapposition were systematically investigated and a comparison made between two commercially available stents (Omega and Biomatrix). Narrower strut spacing led to larger areas of adverse low WSS and high WSSG but these effects were mitigated when strut size was reduced, particularly for WSSG. Local hemodynamics worsened with luminal protrusion of the stent and with stent malapposition, adverse high WSS and WSSG were identified around peak flow and throughout the cardiac cycle respectively. For the Biomatrix stent, the adverse effect of thicker struts was mitigated by greater strut spacing, radial cell offset and flow-aligned struts. In conclusion, adverse hemodynamic effects of specific design features (such as strut size and narrow spacing) can be mitigated when combined with other hemodynamically beneficial design features but increased luminal protrusion can worsen the stent's hemodynamic profile significantly.

Keywords: Computational fluid dynamics (CFD); Coronary artery disease; Hemodynamics; Stent; Stent design; Wall shear stress (WSS).

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Coronary Circulation*
  • Coronary Vessels / physiopathology*
  • Female
  • Hemodynamics*
  • Humans
  • Middle Aged
  • Models, Cardiovascular*
  • Prosthesis Design*
  • Stents*