High-resolution variable-density 3D cones coronary MRA

Magn Reson Med. 2015 Sep;74(3):614-21. doi: 10.1002/mrm.25803. Epub 2015 Jul 14.

Abstract

Purpose: To improve the spatial/temporal resolution of whole-heart coronary MR angiography by developing a variable-density (VD) 3D cones acquisition suitable for image reconstruction with parallel imaging and compressed sensing techniques.

Methods: A VD 3D cones trajectory design incorporates both radial and spiral trajectory undersampling techniques to achieve higher resolution. This design is used to generate a VD 3D cones trajectory with 0.8 mm/66 ms isotropic spatial/temporal resolution, using a similar number of readouts as our previous fully sampled cones trajectory (1.2 mm/100 ms). Scans of volunteers and patients are performed to evaluate the performance of the VD trajectory, using non-Cartesian L1 -ESPIRiT for high-resolution image reconstruction.

Results: With gridding reconstruction, the high-resolution scans experience an expected drop in signal-to-noise and contrast-to-noise ratios, but with L1 -ESPIRiT, the apparent noise is substantially reduced. Compared with 1.2 mm images, in each volunteer, the L1 -ESPIRiT 0.8 mm images exhibit higher vessel sharpness values in the right and left anterior descending arteries.

Conclusion: Coronary MR angiography with isotropic submillimeter spatial resolution and high temporal resolution can be performed with VD 3D cones to improve the depiction of coronary arteries.

Keywords: 3D cones; coronary imaging; parallel imaging; trajectory design.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Coronary Angiography / methods*
  • Female
  • Humans
  • Imaging, Three-Dimensional / methods*
  • Magnetic Resonance Angiography / methods*
  • Male
  • Middle Aged
  • Phantoms, Imaging
  • Signal-To-Noise Ratio