Tracked "Pick-Up" Ultrasound for Robot-Assisted Minimally Invasive Surgery

IEEE Trans Biomed Eng. 2016 Feb;63(2):260-8. doi: 10.1109/TBME.2015.2453173. Epub 2015 Jul 7.

Abstract

Goal: We present a novel "pick-up" ultrasound transducer for intraabdominal robot-assisted minimally invasive surgery. Such a "pick-up" ultrasound transducer is inserted through an abdominal incision at the beginning of the procedure and remains in the abdominal cavity throughout, eliminating the need for a dedicated port or a patient bedside surgical assistant. The transducer has a handle that can be grasped in a repeatable manner using a da Vinci Prograsp tool, allowing the transducer to be accurately manipulated by the surgeon using the da Vinci Robot. This is one way to enable 3-D tracking of the transducer, and, thus, mapping of the vasculature. The 3-D vascular images can be used to register preoperative CT to intraoperative camera images.

Methods: To demonstrate the feasibility of the approach, we use an ultrasound vessel phantom to register a CT surface model to extracted ultrasound vessel models. The 3-D vascular phantom images are generated by segmenting B-mode images and tracking the pick-up ultrasound transducer with the da Vinci kinematics, internal electromagnetic sensor, or visible fiducials suitable for camera tracking.

Results: Reconstruction results using da Vinci kinematics for tracking give a target registration error of 5.4 ± 1.7 mm.

MeSH terms

  • Equipment Design
  • Feasibility Studies
  • Humans
  • Imaging, Three-Dimensional
  • Minimally Invasive Surgical Procedures / instrumentation
  • Minimally Invasive Surgical Procedures / methods*
  • Robotic Surgical Procedures / instrumentation
  • Robotic Surgical Procedures / methods*
  • Tomography, X-Ray Computed
  • Ultrasonography, Interventional / methods*