Quantification and comparison of visibility and image artifacts of a new liquid fiducial marker in a lung phantom for image-guided radiation therapy

Med Phys. 2015 Jun;42(6):2818-26. doi: 10.1118/1.4919616.

Abstract

Purpose: A new biodegradable liquid fiducial marker was devised to allow for easy insertion in lung tumors using thin needles. The purpose of this study was to evaluate the visibility of the liquid fiducial markers for image-guided radiation therapy and compare to existing solid fiducial markers and to one existing liquid fiducial marker currently commercially available.

Methods: Fiducial marker visibility was quantified in terms of contrast to noise ratio (CNR) on planar kilovoltage x-ray images in a thorax phantom for different concentrations of the radio-opaque component of the new liquid fiducial marker, four solid fiducial markers, and one existing liquid fiducial marker. Additionally, the image artifacts produced on computer tomography (CT) and cone-beam CT (CBCT) of all fiducial markers were quantified.

Results: The authors found that the new liquid fiducial marker with the highest concentration of the radio-opaque component had a CNR > 2.05 for 62/63 exposures, which compared favorably to the existing solid fiducial markers and to the existing liquid fiducial marker evaluated. On CT and CBCT, the new liquid fiducial marker with the highest concentration produced lower streaking index artifact (30 and 14, respectively) than the solid gold markers (113 and 20, respectively) and the existing liquid fiducial marker (39 and 20, respectively). The size of the image artifact was larger for all of the liquid fiducial markers compared to the solid fiducial markers because of their larger physical size.

Conclusions: The visibility and the image artifacts produced by the new liquid fiducial markers were comparable to existing solid fiducial markers and the existing liquid fiducial marker. The authors conclude that the new liquid fiducial marker represents an alternative to the fiducial markers tested.

Publication types

  • Comparative Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Artifacts*
  • Cone-Beam Computed Tomography
  • Fiducial Markers*
  • Lung / diagnostic imaging*
  • Lung / radiation effects*
  • Particle Accelerators
  • Phantoms, Imaging*
  • Radiotherapy, Image-Guided / instrumentation*
  • Radiotherapy, Image-Guided / standards*
  • Signal-To-Noise Ratio