Breaking bad IMRT QA practice

J Appl Clin Med Phys. 2015 May 8;16(3):5242. doi: 10.1120/jacmp.v16i3.5242.

Abstract

Agreement between planned and delivered dose distributions for patient-specific quality assurance in routine clinical practice is predominantly assessed utilizing the gamma index method. Several reports, however, fundamentally question current IMRT QA practice due to poor sensitivity and specificity of the standard gamma index implementation. An alternative is to employ dose volume histogram (DVH)-based metrics. An analysis based on the AAPM TG 53 and ESTRO booklet No.7 recommendations for QA of treatment planning systems reveals deficiencies in the current "state of the art" IMRT QA, no matter which metric is selected. The set of IMRT benchmark plans were planned, delivered, and analyzed by following guidance of the AAPM TG 119 report. The recommended point dose and planar dose measurements were obtained using a PinPoint ionization chamber, EDR2 radiographic film, and a 2D ionization chamber array. Gamma index criteria {3% (global), 3 mm} and {3% (local), 3 mm} were used to assess the agreement between calculated and delivered planar dose distributions. Next, the AAPM TG 53 and ESTRO booklet No.7 recommendations were followed by dividing dose distributions into four distinct regions: the high-dose (HD) or umbra region, the high-gradient (HG) or penumbra region, the medium-dose (MD) region, and the low-dose (LD) region. A different gamma passing criteria was defined for each region, i.e., a "divide and conquer" (D&C) gamma method was utilized. The D&C gamma analysis was subsequently tested on 50 datasets of previously treated patients. Measured point dose and planar dose distributions compared favorably with TG 119 benchmark data. For all complex tests, the percentage of points passing the conventional {3% (global), 3 mm} gamma criteria was 97.2% ± 3.2% and 95.7% ± 1.2% for film and 2D ionization chamber array, respectively. By dividing 2D ionization chamber array dose measurements into regions and applying 3mm isodose point distance and variable local point dose difference criteria of 7%, 15%, 25%, and 40% for HD, HG, MD, and LD regions, respectively, a 93.4% ± 2.3% gamma passing rate was obtained. Identical criteria applied using the D&C gamma technique on 50 clinical treatment plans resulted in a 97.9% ± 2.3% gamma passing score. Based on the TG 119 standard, meeting or exceeding the benchmark results would indicate an exemplary IMRT QA program. In contrast to TG 119 analysis, a different scrutiny on the same set of data, which follows the AAPM TG 53 and ESTRO booklet No.7 guidelines, reveals a much poorer agreement between calculated and measured dose distributions with large local point dose differences within different dose regions. This observation may challenge the conventional wisdom that an IMRT QA program is producing acceptable results.

MeSH terms

  • Algorithms*
  • Internationality
  • Practice Guidelines as Topic*
  • Quality Assurance, Health Care / standards*
  • Radiometry / standards*
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / standards*
  • Radiotherapy, Conformal / standards*
  • Reproducibility of Results
  • Sensitivity and Specificity