Automatic Cardiac Self-Gating of Small-Animal PET Data

Mol Imaging Biol. 2016 Feb;18(1):109-16. doi: 10.1007/s11307-015-0868-y.

Abstract

Purpose: The cardiac gating signal (CGS) in positron emission tomography (PET) studies is usually obtained from an electrocardiography (ECG) monitor. In this work, we propose a method to obtain the CGS in small-animal PET using the acquired list-mode data without using any hardware or end-user input.

Procedures: The CGS was obtained from the number of coincidences over time acquired in the lines-of-response connected with the cardiac region. This region is identified in the image as its value changes with frequencies in the range of 3 to 12 Hz. The procedure was tested in a study with 29 Wistar rats and 6 mice injected with 2-deoxy-2-[(18)F]fluoro-D-glucose, which underwent a 45-min single-bed list-mode PET scan of the heart syncronized with an ECG. The estimated signals and the reconstructed images using eight-gated frames were compared with the ones obtained using the ECG signal from the monitor.

Results: The differences of the PET-based CGS with respect to the ECG relative to the duration of the heartbeats were 5.6 % in rats and 11.0 % in mice. The reconstructed gated images obtained from the proposed method do not differ qualitatively with respect to the ones obtained with the ECG. The quantitative analysis of both set of images were performed measuring the volume of the left ventricle (LV) of the rats in the end-of-systole and end-of-diastole phase. The differences found in these parameters between both methods were below 12.1 % in the ESV and 9.3 % in the EDV with a 95 % confidence interval, which are comparable to the accuracy (7 %) of the method used for segmenting the LV.

Conclusion: The proposed method is able to provide a valid and accurate CGS in small-animal PET list-mode data.

Keywords: Cardiac gating; List-mode; PET; Self-gating; Small animal.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Automation
  • Cardiac-Gated Imaging Techniques / methods*
  • Diastole
  • Heart Rate
  • Mice
  • Positron-Emission Tomography / methods*
  • Rats, Wistar
  • Signal-To-Noise Ratio
  • Systole