Quantitative ultrasound molecular imaging

Ultrasound Med Biol. 2015 Sep;41(9):2478-96. doi: 10.1016/j.ultrasmedbio.2015.04.011. Epub 2015 Jun 1.

Abstract

Ultrasound molecular imaging using targeting microbubbles is predominantly a semi-quantitative tool, thus limiting its potential diagnostic power and clinical applications. In the work described here, we developed a novel method for acoustic quantification of molecular expression. E-Selectin expression in the mouse heart was induced by lipopolysaccharide. Real-time ultrasound imaging of E-selectin expression in the heart was performed using E-selectin-targeting microbubbles and a clinical ultrasound scanner in contrast pulse sequencing mode at 14 MHz, with a mechanical index of 0.22-0.26. The level of E-selectin expression was quantified using a novel time-signal intensity curve analytical method based on bubble elimination, which consisted of curve-fitting the bi-exponential equation [Formula: see text] to the elimination phase of the myocardial time-signal intensity curve. Ar and Af represent the maximum signal intensities of the retained and freely circulating bubbles in the myocardium, respectively; λr and λf represent the elimination rate constants of the retained and freely circulating bubbles in the myocardium, respectively. Ar correlated strongly with the level of E-selectin expression (|r|>0.8), determined using reverse transcriptase real-time quantitative polymerase chain reaction, and the duration of post-lipopolysaccharide treatment-both linearly related to cell surface E-selectin protein (actual bubble target) concentration in the expression range imaged. Compared with a conventional acoustic quantification method (which used retained bubble signal intensity at 20 min post-bubble injection), this new approach exhibited greater dynamic range and sensitivity and was able to simultaneously quantify other useful characteristics (e.g., the microbubble half-life). In conclusion, quantitative determination of the level of molecular expression is feasible acoustically using a time-signal intensity curve analytical method based on bubble elimination.

Keywords: Contrast agent; Contrast echocardiography; E-Selectin; Echocardiography; Microbubble elimination; Molecular imaging; Quantification; Targeted microbubbles; Time–signal intensity curve; Ultrasound imaging.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Antibodies, Monoclonal / pharmacokinetics*
  • E-Selectin / metabolism*
  • Echocardiography / methods*
  • Image Interpretation, Computer-Assisted / methods
  • Mice
  • Microbubbles*
  • Molecular Imaging / methods*
  • Myocardium / metabolism*
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Antibodies, Monoclonal
  • E-Selectin