High-Pitch Dual-Source Computed Tomography Renal Angiography Comparison With Conventional Low-Pitch Computed Tomography Angiography: Image Quality, Contrast Medium Volume, and Radiation Dose

J Comput Assist Tomogr. 2015 Sep-Oct;39(5):737-40. doi: 10.1097/RCT.0000000000000268.

Abstract

Objective: The objective of this study was to compare the image quality (IQ), contrast medium (CM) volume, and radiation dose of the high-pitch renal computed tomography angiography (CTA) with low-pitch protocol.

Methods: Fifty patients underwent renal CTA on a dual-source 128-slice scanner via a high-pitch mode (pitch = 2.05) with 0.5-mL/kg CM injection, whereas 50 patients were also scanned on the same scanner with low-pitch (pitch = 0.6) and 1.0 mL/kg CM injection. Subjective IQ was evaluated. Objective IQ was determined by the signal-to-noise ratio and contrast-to-noise ratio. Effective radiation dose was also evaluated.

Results: The contrast-to-noise ratio and signal-to-noise ratio values as well as the IQ scores between the 2 groups had no significant differences (P > 0.05). The effective radiation dose of the high-pitch group was significantly lower (P < 0.05).

Conclusions: High-pitch scan can provide similar subjective and objective IQ compared with low-pitch protocol for renal CTA, whereas CM volume and radiation exposure were significantly reduced.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Angiography / methods*
  • Female
  • Humans
  • Kidney / blood supply
  • Kidney / diagnostic imaging
  • Male
  • Middle Aged
  • Radiation Dosage*
  • Radiographic Image Interpretation, Computer-Assisted / methods*
  • Radiography, Dual-Energy Scanned Projection / methods*
  • Renal Artery / diagnostic imaging*
  • Reproducibility of Results
  • Signal-To-Noise Ratio
  • Tomography, X-Ray Computed / methods*
  • Young Adult