Influence of temporal regularization and radial undersampling factor on compressed sensing reconstruction in dynamic contrast enhanced MRI of the breast

J Magn Reson Imaging. 2016 Jan;43(1):261-9. doi: 10.1002/jmri.24961. Epub 2015 Jun 1.

Abstract

Background: To evaluate the influence of temporal sparsity regularization and radial undersampling on compressed sensing reconstruction of dynamic contrast-enhanced (DCE) MRI, using the iterative Golden-angle RAdial Sparse Parallel (iGRASP) MRI technique in the setting of breast cancer evaluation.

Methods: DCE-MRI examinations of the breast (n = 7) were conducted using iGRASP at 3 Tesla. Images were reconstructed with five different radial undersampling schemes corresponding to temporal resolutions between 2 and 13.4 s/frame and with four different weights for temporal sparsity regularization (λ = 0.1, 0.5, 2, and 6 times of noise level). Image similarity to time-averaged reference images was assessed by two breast radiologists and using quantitative metrics. Temporal similarity was measured in terms of wash-in slope and contrast kinetic model parameters.

Results: iGRASP images reconstructed with λ = 2 and 5.1 s/frame had significantly (P < 0.05) higher similarity to time-averaged reference images than the images with other reconstruction parameters (mutual information (MI) >5%), in agreement with the assessment of two breast radiologists. Higher undersampling (temporal resolution < 5.1 s/frame) required stronger temporal sparsity regularization (λ ≥ 2) to remove streaking aliasing artifacts (MI > 23% between λ = 2 and 0.5). The difference between the kinetic-model transfer rates of benign and malignant groups decreased as temporal resolution decreased (82% between 2 and 13.4 s/frame).

Conclusion: This study demonstrates objective spatial and temporal similarity measures can be used to assess the influence of sparsity constraint and undersampling in compressed sensing DCE-MRI and also shows that the iGRASP method provides the flexibility of optimizing these reconstruction parameters in the postprocessing stage using the same acquired data.

Keywords: DCE-MRI; breast cancer; compressed sensing; golden-angle; iGRASP; parallel imaging; radial undersampling; temporal sparsity.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Algorithms
  • Artifacts*
  • Breast Neoplasms / pathology*
  • Contrast Media
  • Data Compression / methods*
  • Female
  • Gadolinium DTPA
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods
  • Imaging, Three-Dimensional / methods
  • Magnetic Resonance Imaging / methods*
  • Reproducibility of Results
  • Sample Size
  • Sensitivity and Specificity
  • Spatio-Temporal Analysis