General approach for electrochemical detection of persistent pharmaceutical micropollutants: Application to acetaminophen

Biosens Bioelectron. 2015 Oct 15:72:205-10. doi: 10.1016/j.bios.2015.05.010. Epub 2015 May 6.

Abstract

We propose in this work a general and versatile methodology for electrochemical monitoring of persistent pharmaceutical micropollutants. The system presented is based on an electroactive and electropolymerized hapten (mimetic molecule of the pollutant to be detected) and a specific antibody that competitively binds either the hapten or the pollutant. The current delivered by the device depends on this competitive equilibrium and therefore on the pollutant's concentration. The determination of the pharmaceutical product operates within minutes, using square wave voltammetry without labeling or addition of a reactant in solution; the competitive hapten/antibody transduction produces a "signal-on" (a current increase). Applied to acetaminophen, this electrochemical immunosensor presents a very low detection limit of ca. 10 pM, (S/N=3) and a very high selectivity towards structural analogs (aspirin, BPA, and piperazine) even in a mixture.

Keywords: Acetaminophen; Competitive immunoassay; Electrochemical immunosensor; Hapten; Label-free; Reagentless.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Acetaminophen / analysis*
  • Biosensing Techniques / methods
  • Click Chemistry
  • Electrochemical Techniques / methods*
  • Haptens / chemistry
  • Immunoassay / methods
  • Limit of Detection
  • Water Pollutants, Chemical / analysis*