[Uncertainties in the localization of radiation sources in brachytherapy]

Strahlenther Onkol. 1989 Nov;165(11):807-12.
[Article in German]

Abstract

While investigating ways to solve the problem of locating radiation sources applied in brachytherapy a device was constructed which enables a precise assessment of their or their applicators' position from two only roughly positioned orthogonal radiographs. The overall accuracy (including exposure and evaluation) achieved is +/- 1 mm. The constructional elements and the evaluation algorithm are described and the influence of the location uncertainty on the uncertainty of the dose applied are discussed.

Publication types

  • Comparative Study
  • English Abstract

MeSH terms

  • Algorithms
  • Brachytherapy / instrumentation*
  • Humans
  • Models, Biological
  • Radiotherapy Dosage*