Low-Cost Disposable Cartridge for Performing a White Blood Cell Count and Partial Differential at the Point-of-Care

Health Innov Point Care Conf. 2014 Oct:2014:10-13. doi: 10.1109/HIC.2014.7038862.

Abstract

Being able to perform a white blood cell (WBC) count and differential is a crucial laboratory test for basic diagnostic practices. In this paper, we demonstrate proof of concept results for a disposable cartridge that could be used to perform a WBC count and 3-part differential at the point-of-care. The cartridge is composed of a glass slide, a layer of transfer tape, and a glass cover slip and incorporates acridine orange for cell staining and sub-type differentiation; the stained blood is then imaged, and image analysis techniques return a WBC count and 3-part differential. The cartridge was tested on a laboratory microscope with 3 normal samples, and had promising results with 85.7% of images resulting in a WBC count with ±15% of the true value. Further, the 3-part differential concentrations determined using the disposable cartridge had strong correlations with the true concentrations (R2 values of 0.9986, 0.9421, and 0.6942 for granulocytes, lymphocytes, and monocytes, respectively). Preliminary designs for a low-cost, portable microscope have been created and are currently being prototyped.