Immunity to Laser Power Variation in a DFB Diode Laser Based Optical Gas Sensor Using a Division Process

Sensors (Basel). 2015 Apr 22;15(4):9582-91. doi: 10.3390/s150409582.

Abstract

The division process used in a DFB diode laser-based optical gas sensor was studied to improve the immunity to laser power variation. Residual amplitude modulation (RAM) in wavelength modulation spectroscopy (WMS) detection was eliminated by intensity normalization using a division process. As a result the detected harmonic signals showed a significant improvement in line shape. For the first harmonic (1f) signal, Bias was improved from 38.7% to 1.2%; Baseline Difference was improved from 2.7% to 0.69% and Asymmetry was improved from 15.4% to 0.22%. For the second harmonic (2f) signal, the Asymmetry Coefficient was improved from 103% to 5.1%. Moreover the division process can further suppress the influence of unstable laser power. As a result, for the 1f signal, stable detection with a variation coefficient of 0.59% was obtained over a wide dynamic range (0.38-8.1 mW). For the 2f signal, stable detection with a variation coefficient of 0.53% was obtained from 0.64 mW to 8.27 mW. The test results showed a good agreement with the theoretical analysis and the proposed method has considerable potential application in gas sensing.