A prospective observational study comparing a non-operator dependent automatic PWV analyser to pulse pressure, in assessing arterial stiffness in hemodialysis

BMC Nephrol. 2015 Apr 23:16:62. doi: 10.1186/s12882-015-0058-9.

Abstract

Background: Chronic kidney disease (CKD) accelerates vascular stiffening related to age. Arterial stiffness may be evaluated measuring the carotid-femoral pulse wave velocity (PWV) or more simply, as recommended by KDOQI, monitoring pulse pressure (PP). Both correlate to survival and incidence of cardiovascular disease. PWV can also be estimated on the brachial artery using a Mobil-O-Graph; a non-operator dependent automatic device. The aim was to analyse whether, in a dialysis population, PWV obtained by Mobil-O-Graph (MogPWV) is more sensitive for vascular aging than PP.

Methods: A cohort of 143 patients from 4 dialysis units has been followed measuring MogPWV and PP every 3 to 6 months and compared to a control group with the same risk factors but an eGFR > 30 ml/min.

Results: MogPWV contrarily to PP did discriminate the dialysis population from the control group. The mean difference translated in age between the two populations was 8.4 years. The increase in MogPWV, as a function of age, was more rapid in the dialysis group. 13.3% of the dialysis patients but only 3.0% of the control group were outliers for MogPWV. The mortality rate (16 out of 143) was similar in outliers and inliers (7.4 and 8.0%/year). Stratifying patients according to MogPWV, a significant difference in survival was seen. A high parathormone (PTH) and to be dialysed for a hypertensive nephropathy were associated to a higher baseline MogPWV.

Conclusions: Assessing PWV on the brachial artery using a Mobil-O-Graph is a valid and simple alternative, which, in the dialysis population, is more sensitive for vascular aging than PP. As demonstrated in previous studies PWV correlates to mortality. Among specific CKD risk factors only PTH is associated with a higher baseline PWV.

Trial registration: ClinicalTrials.gov Identifier: NCT02327962.

Publication types

  • Observational Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Automation
  • Blood Pressure / physiology*
  • Brachial Artery / diagnostic imaging
  • Brachial Artery / physiopathology*
  • Cohort Studies
  • Female
  • Humans
  • Kidney Failure, Chronic / physiopathology*
  • Kidney Failure, Chronic / therapy
  • Male
  • Middle Aged
  • Prospective Studies
  • Pulse Wave Analysis / instrumentation
  • Pulse Wave Analysis / methods*
  • Renal Dialysis
  • Sensitivity and Specificity
  • Sphygmomanometers
  • Ultrasonography
  • Vascular Stiffness / physiology*

Associated data

  • ClinicalTrials.gov/NCT02327962