Real Time Three-Dimensional Echocardiographic Evaluations of Fetal Left Ventricular Stroke Volume, Mass, and Myocardial Strain: In Vitro and In Vivo Experimental Study

Echocardiography. 2015 Nov;32(11):1697-706. doi: 10.1111/echo.12939. Epub 2015 Apr 11.

Abstract

Background: Left ventricular stroke volume, mass, and myocardial strain are valuable indicators of fetal heart function. This study investigated the feasibility of nongated real time three-dimensional echocardiography (RT3DE) to determine fetal stroke volume (SV), left ventricular mass (LVM), and myocardial strain under different conditions.

Methods: To evaluate fetal hearts, fetal-sized rabbit hearts were used in this study. The in vitro portion of this study was carried out using a balloon inserted into the LV of eight fresh rabbit hearts and driven by a calibrated pulsatile pump. RT3DE volumes were obtained at various pump-set SVs. The in vivo experiments in this study were performed on open-chest rabbits. RT3DE volumes were acquired at the following conditions: baseline, simulated hypervolemia, inferior vena cava (IVC) ligation, and ascending aorta (AAO) ligation. Displacement values and sonomicrometry data were used as references for RT3DE-derived SV, LVM, longitudinal strain (LS), and circumferential strain (CS).

Results: Excellent correlations between RT3DE-derived values and reference values were demonstrated and accompanied by high coefficients of determination (R(2) ) for both in vitro and in vivo studies for SV, LVM, LS, and CS (in vitro: SV: R(2) = 0.98; LVM: R(2) = 0.97; LS: R(2) = 0.87, CS: R(2) = 0.80; in vivo: SV: R(2) = 0.92; LVM: R(2) = 0.98; LS: in vivo: R(2) = 0.84; CS: in vivo: R(2) = 0.76; all P < 0.05).

Conclusions: RT3DE is capable of quantifying the SV, LVM, and myocardial strain of fetal-sized hearts under different conditions. This nongated RT3DE may aid the evaluation of fetal cardiac function, providing a superior understanding of the progress of fetal heart disorders.

Keywords: fetal echocardiography; left ventricular mass, strain; real time imaging technique.

MeSH terms

  • Animals
  • Disease Models, Animal
  • Echocardiography, Three-Dimensional*
  • Female
  • Fetal Heart / diagnostic imaging*
  • Fetal Heart / physiopathology*
  • Heart Ventricles / diagnostic imaging*
  • Heart Ventricles / pathology
  • Heart Ventricles / physiopathology*
  • In Vitro Techniques
  • Organ Size
  • Rabbits
  • Stroke Volume / physiology*
  • Ventricular Dysfunction, Left / diagnostic imaging
  • Ventricular Dysfunction, Left / pathology
  • Ventricular Dysfunction, Left / physiopathology