Performance evaluation of two computed radiography systems and patient dose in pelvic examination

Radiat Prot Dosimetry. 2015 Jul;165(1-4):392-6. doi: 10.1093/rpd/ncv125. Epub 2015 Apr 1.

Abstract

This study was carried out to evaluate the performance of two computed radiography (CR) units. These evaluations became necessary following the introduction of CR systems in Sudan. Evaluation of the CR systems was performed using physical image quality parameters: signal transfer property, modulation transfer function, normalised noise power spectrum, detective quantum efficiency and the subjective contrast detail detectability. Patient dose was measured in terms of entrance surface air kerma estimated from tube output and exposure factors for 100 patients who had undergone pelvic X-ray examinations. Fuji computed radiography velocity system with columnar screen dose results was much lower than those using CR975 system with granular screen. Patient doses delivered by both systems were within the international diagnostic reference levels.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Humans
  • Pelvis / diagnostic imaging*
  • Radiation Dosage
  • Radiographic Image Enhancement / instrumentation*
  • Radiographic Image Enhancement / methods*
  • Radiometry
  • Reproducibility of Results
  • Signal-To-Noise Ratio