Associations of Resonance Frequency Analysis with Dynamic Mechanical Analysis of Dental Implant Systems

Clin Implant Dent Relat Res. 2016 Apr;18(2):332-41. doi: 10.1111/cid.12319. Epub 2015 Mar 23.

Abstract

Background: Resonance frequency analysis (RFA) has been introduced as a noninvasive method to clinically estimate the stability of dental implant systems.

Purpose: The objective of this study was to examine whether implant stability quotient (ISQ) values of RFA can account for mechanical stability of the dental implant system, which is assessed using dynamic mechanical analysis (DMA).

Materials and methods: Fifty-seven screw-type titanium dental implants were placed in artificial polyurethane foams with seven different thicknesses (3.5 to 12 mm) and eight edentulous mandibles of human cadavers (four men and four women, 79.11 ± 13.48 years). After the ISQ values, insertion torque, and static stiffness of each implant system were measured, the DMA was performed to assess dynamic stiffness and viscoelastic tan δ.

Results: The ISQ value had strong positive correlations with thickness, insertion torque, static and dynamic stiffness, and a negative correlation with tan δ of implant systems in artificial bone blocks (r = 0.769 to 0.992, p < .043). However, the ISQ value was correlated with only the insertion torque of implant systems in human mandibles (p < .049).

Conclusion: The ISQ values could reflect mechanical stability of the dental implant system under the controlled condition of homogeneous density in simple dimensions.

Keywords: dental implant system; dynamic mechanical analysis; implant stability quotient; insertion torque; resonance frequency analysis.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Aged, 80 and over
  • Bone Density
  • Dental Implantation, Endosseous*
  • Dental Implants
  • Dental Prosthesis Retention*
  • Female
  • Humans
  • Male
  • Middle Aged
  • Resonance Frequency Analysis*
  • Torque
  • Vibration

Substances

  • Dental Implants