Evaluation of an instrument-assisted dynamic prosthetic alignment technique for individuals with transtibial amputation

Prosthet Orthot Int. 2016 Aug;40(4):475-83. doi: 10.1177/0309364615574161. Epub 2015 Mar 11.

Abstract

Background: A prosthesis that is not optimally aligned can adversely influence the rehabilitation and health of the amputee. Very few studies to date evaluate the effectiveness and utility of instrument-assisted alignment techniques in clinical practice.

Objectives: To compare an instrument-assisted dynamic alignment technique (Compas(™)) to conventional methods.

Study design: In a crossover study design, dynamic prosthetic alignments were provided to nine individuals with unilateral transtibial amputations to compare conventional and instrument-assisted alignment techniques.

Methods: The instrument-assisted technique involved a commercially available force and torque sensing dynamic alignment system (Compas). Cadence, pelvic accelerations, and socket moments were assessed. A custom questionnaire was used to gather user perceptions.

Results: No differences between alignment techniques were found in global gait measures including cadence and pelvic accelerations. No significant alignment differences were achieved by examination of angular changes between the socket and foot; however, significantly higher below-the-socket moments were found with the instrument-assisted technique. From the questionnaire, six amputees had no preference, while three preferred the conventional alignment.

Conclusion: The use of Compas appears to produce similar alignment results as conventional techniques, although with slightly higher moments at the socket.

Clinical relevance: This study provides new information about the clinical utilization of instrument-assisted prosthetic alignment techniques for individuals with transtibial amputation.

Keywords: Testing of prosthetic and orthotic components; biomechanics of prosthetic/orthotic devices; gait analysis; prosthetic alignment; rehabilitation of prostheses users; transtibial amputation.

MeSH terms

  • Adult
  • Aged, 80 and over
  • Amputation, Surgical*
  • Artificial Limbs*
  • Cross-Over Studies
  • Female
  • Gait / physiology*
  • Humans
  • Lower Extremity
  • Male
  • Middle Aged
  • Prosthesis Design
  • Prosthesis Fitting*
  • Range of Motion, Articular / physiology
  • Tibia
  • Young Adult