Clinical comparison of QUANTA Flash dsDNA chemiluminescent immunoassay with four current assays for the detection of anti-dsDNA autoantibodies

J Immunol Res. 2015:2015:902821. doi: 10.1155/2015/902821. Epub 2015 Jan 5.

Abstract

Introduction: The objective of the present study was to compare QUANTA Flash dsDNA, a chemiluminescent immunoassay (CIA) on the BIO-FLASH, a rapid-response chemiluminescent analyzer, to three other anti-dsDNA antibody assays and to Crithidia luciliae indirect immunofluorescence test (CLIFT).

Methods: In the first part of the study, 161 samples, 61 from patients suffering from systemic lupus erythematosus (SLE) and 100 from a disease control group, were tested by QUANTA Flash dsDNA CIA, QUANTA Lite dsDNA SC ELISA, BioPlex 2200 multiplex flow immunoassay (MFI), ImmuLisa dsDNA ELISA, and NOVA Lite CLIFT. A second cohort of 69 SLE patients was then tested by QUANTA Flash dsDNA and CLIFT to expand the study.

Results: The overall qualitative agreements varied between 77.0% (NOVA Lite CLIFT versus QUANTA Lite) and 89.4% (ImmuLisa versus NOVA Lite CLIFT). The clinical sensitivities for the anti-dsDNA antibody tests varied from 8.2% (NOVA Lite CLIFT) to 54.1% (QUANTA Lite), while the clinical specificities varied from 88.0% (BioPlex 2200) to 100.0% (NOVA Lite CLIFT). Good correlation was found between QUANTA Flash dsDNA and NOVA Lite CLIFT.

Conclusion: Significant variations among dsDNA methods were observed. QUANTA Flash dsDNA provides a good combination of sensitivity and specificity for the diagnosis of SLE and good agreement to CLIFT.

Publication types

  • Comparative Study

MeSH terms

  • Antibodies, Antinuclear / blood*
  • Case-Control Studies
  • Cluster Analysis
  • Fluorescent Antibody Technique* / methods
  • Fluorescent Antibody Technique* / standards
  • Humans
  • Lupus Erythematosus, Systemic / blood*
  • Lupus Erythematosus, Systemic / diagnosis*
  • ROC Curve
  • Reproducibility of Results
  • Sensitivity and Specificity

Substances

  • Antibodies, Antinuclear