In vivo quantification of renal function in mice using clinical gamma cameras

Phys Med. 2015 May;31(3):242-7. doi: 10.1016/j.ejmp.2015.01.013. Epub 2015 Feb 26.

Abstract

Introduction: In preclinical research, the growing number of transgenic models has led to the need for renal-function studies in mice. Many efforts have been made to develop dedicated SPECT systems for rodents, but their availability is limited due to high capital costs. The aim of this work is to demonstrate the feasibility of mouse renal imaging by using an inexpensive alternative based on clinical gamma-cameras.

Methods: A healthy mouse was scanned 3 h after injection of 6 mCi of Dimercaptosuccinic acid (DMSA) labeled with 99mTc by using a single-head gamma-camera in conjunction with a dedicated pinhole collimator. List-mode data were binned to emulate multiple injections of 1 mCi, 0.1 mCi and 0.01 mCi of 99mTc-DMSA and 6-min ventral and dorsal planar images were acquired and SPECT imaging (60 projection images acquired over 60 min) was performed. An optimization of the protocols in terms of injected activity, time scan, renal cortex uniformity and cortex-to-pelvis contrast was carried out.

Results: The appropriate protocols were an injected activity of 0.6 mCi, combined with duration of scanning of 1 min for planar and 60 min for SPECT imaging. Our results were validated through the relative quantification of renal function, which showed that both kidneys contributed equally to the total function. They showed that functional structures of the mouse kidneys can be visually distinguished as easily as in human studies.

Conclusions: Our findings showed the feasibility of conducting quantitative DMSA SPECT studies of anesthetized mice on clinical gamma cameras.

Keywords: DMSA; Pinhole SPECT; Quantitative imaging; Renal imaging; Small animal imaging.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Calibration
  • Gamma Cameras*
  • Humans
  • Image Processing, Computer-Assisted
  • Injections
  • Kidney Tubules, Proximal / diagnostic imaging*
  • Kidney Tubules, Proximal / physiology*
  • Mechanical Phenomena
  • Mice
  • Succimer
  • Technetium
  • Time Factors
  • Tomography, Emission-Computed, Single-Photon / instrumentation*

Substances

  • Technetium
  • Succimer