Rapid method for the quantification of hydroquinone concentration: chemiluminescent analysis

Luminescence. 2015 Nov;30(7):947-9. doi: 10.1002/bio.2842. Epub 2015 Feb 18.

Abstract

Topical hydroquinone serves as a skin whitener and is usually available in cosmetics or on prescription based on the hydroquinone concentration. Quantification of hydroquinone content therefore becomes an important issue in topical agents. High-performance liquid chromatography (HPLC) is the commonest method for determining hydroquinone content in topical agents, but this method is time-consuming and uses many solvents that can become an environmental issue. We report a rapid method for quantifying hydroquinone content by chemiluminescent analysis. Hydroquinone induces the production of hydrogen peroxide in the presence of basic compounds. Hydrogen peroxide induced by hydroquinone oxidized light-emitting materials such as lucigenin, resulted in the production of ultra-weak chemiluminescence that was detected by a chemiluminescence analyzer. The intensity of the chemiluminescence was found to be proportional to the hydroquinone concentration. We suggest that the rapid (measurement time, 60 s) and virtually solvent-free (solvent volume, <2 mL) chemiluminescent method described here for quantifying hydroquinone content may be an alternative to HPLC analysis.

Keywords: chemiluminescence; hydroquinone; skin whitener; topical.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromatography, High Pressure Liquid
  • Hydrogen Peroxide / chemistry
  • Hydroquinones / analysis*
  • Luminescent Measurements*
  • Molecular Structure
  • Oxidation-Reduction

Substances

  • Hydroquinones
  • Hydrogen Peroxide
  • hydroquinone